Combination With Other Ceiling Supply Units; Flexibility Trolley For Ceiling-Mounted Version - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Assembly and preparation
Product label 2:
C
 Make sure that the keyswitch on the ceiling
supply unit is in position (A). Contact service
personnel if necessary.
WARNING
Risk of injury
If the ceiling supply unit has not been set up
for operation with Perseus or the keyswitch is
in an incorrect setting position, the safety
sensor will be inoperative.
Only operate Perseus with ceiling supply
units that have been set up for Perseus and
have the keyswitch in the correct setting
position. If necessary, contact service
personnel for clarification and specialized
service personnel if the keyswitch is faulty.
48
A
B
Combination with other ceiling supply
units
Perseus can be used with ceiling supply units from
other manufacturers. In this case, a special kit (see
list of accessories) is required. The ceiling supply
unit must be prepared before Perseus can be
mounted.
Dräger will provide the manufacturer of the ceiling
supply unit with the document "Perseus A500
anesthesia workstation - ceiling-mounted version -
interface description for ceiling supply units". The
instructions in this document must be followed
during construction.
Flexibility trolley for ceiling-mounted
version
The ceiling-mounted version of Perseus can be
mounted on a flexibility trolley and can then be
used as a trolley version.
When Perseus is mounted on the flexibility trolley,
the remote control is inoperative and the safety
sensor is disabled.
Information on attaching accessories to Perseus
on the flexibility trolley can be found in the chapter
"Mounting of accessories".
Instructions for use Perseus A500 SW 2.0n