Pressure-Controlled Ventilation - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Pressure-controlled ventilation

PC
Paw
Slope
Flow
PC - CMV
– Pressure-controlled
– Time-controlled
– Machine-triggered
The mandatory breaths are machine-triggered and
are not triggered by the patient.
PC - BIPAP
– Pressure-controlled
– Time-controlled
– Machine-triggered or patient-triggered
– Synchronized inspiration
Instructions for use Perseus A500 SW 2.0n
Pinsp
PEEP
Ti
1/RR
Trigger
∆Psupp
Slope
Trigger
Insp. term.
In PC - BIPAP, the patient can breathe
spontaneously at any time, while the number of
mandatory breaths is predefined. When
synchronization is switched on, the breaths are
adapted to the spontaneous breathing efforts of
the patient. If spontaneous breathing effort by the
patient is detected during the inspiratory trigger
window, a patient-triggered breath will be initiated.
PC - BIPAP / PS (optional)
This mode is similar to PC - BIPAP, except that the
patient's spontaneous breathing at the PEEP level
during the expiratory phase is pressure-supported
with ∆Psupp when outside the trigger window.
Principles of operation
263