Breathing System - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Breathing system

A Water trap with connection for sample line
B Pressure gauge (optional) for indicating the
pressure in the internal breathing system
The pressure in the internal breathing system
is also indicated on the status display, see (G)
on page 31 and (H) on page 32.
WARNING
Pressure gauge indication possibly
inaccurate
Always compare the pressures indicated on
the pressure gauge with those on the status
display. Only use the pressure gauge as the
primary source of information if the pressure
indicator on the status display has failed.
C Collecting channel
D Inspiratory port
Instructions for use Perseus A500 SW 2.0n
A
B
C
D
K
E Bag elbow with circuit plug, e.g., to seal the
Y-piece during an automatic test
F APL valve
G Expiratory port
H Breathing system cover
I
Holder, e.g., for breathing bag hose
J CO
absorber
2
K Anesthetic gas receiving system
E
F
G
J
Overview
H
I
21