Configuration
Vertical tab "Interfaces"
Headline/
Parameter
LAN
DHCP
IP address
Subnet mask
Default gateway
MAC address
COM 1
Protocol
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
COM 2
Protocol
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
USB
USB interface
168
Setting range
On; Off
XXX . XXX . XXX . XXX
XXX . XXX . XXX . XXX
XXX . XXX . XXX . XXX
MEDIBUS.X; None
MEDIBUS.X; None
On; Off
Description
Configure the network.
Make network settings.
When using DHCP, consult with IT
personnel to ensure that Perseus
is always assigned the same IP
address by the DHCP server.
Restart the device after each
change to the network settings.
The network settings are not
affected by a reset to factory set-
tings.
Accept the changes to IP address,
Subnet mask, or Default gateway
with the Apply button. The
changes are only active after the
device has been restarted.
Displays the MAC address.
Configure the COM interfaces.
A baud rate of 19200 or 38400 is
required for transmission of high-
speed data, e.g., for waveforms.
Activate or deactivate the USB
interface.