Device Combinations; Response Times In Event Of Concentration Changes - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data
Breathing system, dry, with filled reusable
CO
absorber and soda lime dust filter
2
MX50115
With breathing circuit for pediatric patients
M27542, inspiratory filter MP01735
With breathing circuit for neonates MP00333,
inspiratory filter MP01735

Response times in event of concentration changes

Typical response times (T
concentration change from 21 Vol% to 100 Vol% at
the following fresh-gas flows:
Test lung for adults (MP02400), breathing circuit
(MP00300), breathing bag 2 L (MP00222)
VT=500 mL, RR=10 /min, I:E=1:2
Test lung for neonates (8410079), breathing cir-
cuit (MP00333), breathing bag 1 L (MP00383)
VT=30 mL, RR=30 /min, I:E=1:1

Device combinations

This device can be operated in combination with
other Dräger devices or with devices from other
manufacturers. Follow the accompanying
documents of the individual devices.
If a device combination is not approved by Dräger,
the safety and the correct functioning of the
individual devices can be compromised. The
operating organization must ensure that the device
combination complies with the applicable editions
of the relevant standards for medical devices.
Device combinations approved by Dräger meet the
requirements of the following standards:
– IEC 60601-1 (electrical safety, mechanical
safety, software)
– IEC 60601-1-2 (EMC)
– IEC 60601-1-8 (alarm systems)
250
) for an oxygen
0..90
Resistance [hPa (cmH
Peak flow in
Man / Spon
use [L/min]
Inspiratory
15
–0.7
2.5
1.1
2 L/min
4 L/min
712 s
174 s
91 s
64 s
Instructions for use Perseus A500 SW 2.0n
O)]
2
Expiratory
3.0
1.7
8 L/min
O
flush
2
32 s
9 s
46 s
7 s