Displaying Alarms; Optical Alarm Signals; Acoustic Alarm Signals - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Displaying alarms

Alarms are signaled optically and acoustically.

Optical alarm signals

On the screen
In the event of an alarm, the system displays the
relevant alarm message in the alarm message
field (A). The parameter field (B) for the alarm-
generating parameter flashes.
C
B
B
B
Up to 8 alarms can be displayed simultaneously in
the alarm message field in the header bar (A). If
more alarms occur, the All alarms button (C) is
displayed in the header bar. Touching this button
opens the dialog window Alarms > Current
alarms with information about all active alarms,
see page 140.
On Dräger Vapor 3000
When the Vapor View option is installed, alarms
concerning the vaporizer are signaled by blinking
illumination on the Dräger Vapor 3000 if their cause
can be remedied by adjusting or refilling the
vaporizer, for example:
– Inspiratory xMAC high
– Filling level of vaporizer low
– Vaporizer open
– xMAC low
Instructions for use Perseus A500 SW 2.0n
A
B
B
B

Acoustic alarm signals

It always is the alarm with the highest priority that
is acoustically signaled. The signal is emitted until
either the cause of the alarm is remedied or the
Audio paused key is pressed.
Depending on the overall alarm situation, it is
possible that the 10-tone sequence for the high
alarm priority is only issued as a 5-tone sequence
due to the coinciding occurrence of alarms.
Regardless of the set alarm volume, the No O
delivery alarm is issued at maximum volume.
Alarms
2
139