Dräger Perseus A500 Instructions For Use Manual page 167

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Headline/
Parameter
Test breathing circuit
Test correct assem-
bly of breathing
hoses and Y-piece
Vertical tab "Logbook"
Headline/
Parameter
A logbook entry with measured values is created
Every
For all high-priority
alarms
For all medium-pri-
ority alarms
Vertical tab "Sound volume"
Headline/
Parameter
Alarm tone volume
Minimum alarm tone
volume
Breathing sound
volume (optional)
Instructions for use Perseus A500 SW 2.0n
Setting range
On; Off
Setting range
1 min; 2 min; 5 min; 10 min; 15 min
On; Off
On; Off
Setting range
10 to 100
40
10 to 100
10
Off; 10 to 100
50
Configuration
Description
Select whether the automatic test
will check that the breathing circuit
is correctly connected.
During this test, a check is made
as to whether the breathing gas
can flow from the inspiratory port
via the Y-piece to the expiratory
port.
If Sample line is connected during
test is set to On, this test is carried
out fully automatically. If Sample
line is connected during test is set
to Off, additional manual checks
are required.
Description
Create additional logbook entries
with measured values.
Create entries regularly.
Create entries for alarms.
Description
Set the alarm tone volume.
Set the minimum volume with
which an alarm tone will be sig-
naled.
Set the breathing sound volume.
167