1 Open the Trends/Data dialog window.
2 Touch the Export tab (A).
Trends/Data
B
B
B

Other settings

Switching the breathing system warmer
on or off
The breathing system warmer should only be
switched off in special situations (e.g., for
intentional reduction of the body temperature of
the patient).
1 Open the System setup dialog window.
2 Touch the Therapy tab (A).
System setup
A
B
3 To switch the warmer on or off, touch the
corresponding button (B).
114
A
3 Touch the desired button (B).
The data will be saved as a ".txt" file.
When switching to Standby mode, the warmer is
reset to the value configured in System setup.
CAUTION
Risk due to faulty or switched-off breathing
system warmer
Increased condensation or accumulation of water
in the breathing system and the hoses may
occur.
Increase fresh-gas flow as required. Remove
condensate from hoses, water traps, and
breathing system regularly. Have service
personnel repair the faulty breathing system
warmer.
CAUTION
Risk of inaccurate measured values
The accuracy of the flow measurement may be
compromised if the breathing system warmer is
switched off.
When the breathing system warmer is switched
off, do not make therapy decisions solely on the
basis of the displayed values for flow and
volume.
Instructions for use Perseus A500 SW 2.0n