Dräger Perseus A500 Instructions For Use Manual page 232

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data
Measuring systems and displays (continued)
Pressure gauge for indicating the pressure in
the internal breathing system
Range
Accuracy
Resolution of displayed value
Volume
Measured volumes are corrected by the
determined breathing hose compliance. As
soon as CO
respiratory phases are
2
detected, measured volumes are also cor-
rected by the suction flow of the integral
patient-gas measurement module and
according to the determined gas composi-
tion.
Tidal volume
Inspiratory
Expiratory
Range
Accuracy
Resolution of displayed value
∆VT
Range
Accuracy
Resolution of displayed value
Minute volume
Total
Mandatory
Spontaneous
Range
Accuracy
Resolution of displayed value
T
0...90
232
–20 to +80 hPa (cmH
2
±5 % of the measured value or ±2 hPa (cmH
(the larger value applies)
5 hPa (cmH
O)
2
VTi
VT
0 to 2500 mL
±8 % of measured value or ±15 mL (the larger
value applies)
1 mL
0 to 2500 mL
±16 % of measured value or ±30 mL (the larger
value applies)
1 mL
MV
MVmand
MVspon
0 to 40 L/min
±8 % of measured value or ±0.2 L/min (the
larger value applies)
0.1 or 0.01 L/min
<45 s (RR ≥6 /min)
<105 s (RR <6 /min)
Instructions for use Perseus A500 SW 2.0n
O)
2