For devices with isolation transformer:
– If necessary, wait 2 minutes for the circuit
breaker to cool.
– Press the key on the circuit breaker back in
or operate the switch on the isolation
transformer.
4 Restore mains power supply.
Emergency O
delivery (electronically controlled gas mixer)
2
WARNING
Risk of increased anesthetic agent delivery
When the emergency O
in use, anesthetic agent continues to be
delivered into the breathing system. When
the emergency O
delivery is used during
2
low-flow anesthesia or minimal-flow
anesthesia, an increased quantity of
anesthetic agent may enter the breathing
system. This may lead to an increased
anesthetic gas concentration.
Carefully monitor the gas mixture.
Instructions for use Perseus A500 SW 2.0n
delivery (Add. O
2
2
B
A
1 Check vaporizer setting.
2 Set the O
2
position.
3 Open the flow control valve (B) on the O
) is
flowmeter and set the desired flow. This O
flow flows through the vaporizer.
Troubleshooting
switch (A) upwards to the Add. O
2
2
2
183