Dräger Perseus A500 Instructions For Use Manual page 119

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

WARNING
Risk due to residual concentrations of
anesthetic agent
Even after flushing the breathing system and
breathing circuit, residual concentrations of
anesthetic agent might not be sufficiently
reduced.
For patients suspected of having malignant
hyperthermia, do not rely solely on the
flushing function.
For further information and recommendations for
therapy settings for patients with suspected
malignant hyperthermia, contact the responsible
national Dräger organization.
CAUTION
Risk of impaired ventilation due to faulty drying
function
The breathing system and breathing hoses may
still retain traces of moisture even after the
flushing and drying functions are performed.
Check the breathing system and breathing hoses
for condensate.
Instructions for use Perseus A500 SW 2.0n
Operation
119