One-Step Recruitment - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Operation
The breath is ended automatically:
– After a maximum of 40 seconds in the Adult
patient category
– After a maximum of 30 seconds in the Ped
patient category
– After a maximum of 5 seconds in the Neo
patient category
Holding or extending the expiration
 Touch the Exp. hold button (D) and keep it
pressed for the desired time.
Perseus will extend the expiration and delay the
next breath.
The expiration is ended automatically:
– After a maximum of 40 seconds in the Adult
patient category
– After a maximum of 30 seconds in the Ped
patient category
– After a maximum of 5 seconds in the Neo
patient category
96

One-step recruitment

This maneuver applies a set pressure for a specific
duration and thus enables, e.g., an extended
breath with adjustable pressure.
Procedures
E
F
1 Using the therapy controls, set the
Pressure (E) and the Duration (F).
2 If necessary, touch the More button (H) and set
a reminder with the therapy control (I).
3 Touch the Start button (G) and confirm.
The rise in pressure from the PEEP level to the set
pressure level and the decrease in pressure at the
end of the maneuver take place at 20 hPa/s.
The maneuver ends automatically after the time
has elapsed. To cancel the maneuver, touch the
Cancel button (J) and confirm. If required, the
PEEP can be adjusted before confirming.
The Paw high alarm limit is checked at the start of
the maneuver. If the alarm limit is too low, it is
changed to a value 5 hPa above the set pressure.
After the maneuver, the alarm limit is reset to its
original value.
Instructions for use Perseus A500 SW 2.0n
I
G
H
J

Advertisement

Table of Contents
loading

Table of Contents