Dräger Perseus A500 Instructions For Use Manual page 158

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
Vertical tab "Vent. 2"
Headline/
Parameter
Start settings for ventilation
Pmax
[mbar]; [hPa];
[cmH
O]
2
Pinsp
[mbar]; [hPa];
[cmH
O]
2
∆Psupp
[mbar]; [hPa];
[cmH
O]
2
Insp. term.
[%PIF]
PEEP
[mbar]; [hPa];
[cmH
O]
2
Slope
[s]
RRmin
[1/min]
I:E
%Tplat
[%]
Trigger
[L/min]
Sync.
158
Setting range
7 to 80
7 to 80
40
30
3 to 80
3 to 80
15
15
Off;
Off;
1 to 80
1 to 80
10
10
5 to 80
5 to 80
25
25
Off;
Off;
2 to 35
2 to 35
3
3
0 to 2
0 to 2
0.2
0.2
Off;
Off;
3 to 25
3 to 25
6
10
1:50 to 50:1
1:50 to 50:1
1:2.0
1:2.0
0 to 60
0 to 60
20
20
0.3 to 15
0.3 to 15
4.0
2.0
On; Off
On; Off
Description
Specify start settings for the venti-
lation.
7 to 80
25
3 to 80
15
Off;
1 to 80
10
5 to 80
25
Off;
2 to 35
3
0 to 2
0.2
Off;
3 to 25
15
1:50 to 50:1
1:1.0
0 to 60
20
0.3 to 15
1.0
On; Off
Instructions for use Perseus A500 SW 2.0n