Technical data
Display of calculated values (continued)
Anesthetic agent
Range
Accuracy
Resolution of displayed value
Waveforms
Sweep speed
Scale
Airway pressure Paw (t)
Flow (t)
Volume (t)
O
(t)
2
CO
(t)
2
Anesthetic agent (t)
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
Loops

Operating characteristics

Mains power supply
Electrical mains connection for the trolley
version
Without isolation transformer
With isolation transformer
238
0 to 999.9 mL liquid
±25 %
0.1 mL
O
concentration
2
Primary anesthetic agent concentration
CO
concentration
2
Airway pressure
Volume (only for loops)
Flow
6.25; 12.5; 25 mm/s
–20 to 80 hPa (cmH
O)
2
–120 to 120 L/min
0 to 2000 mL
0 to 100 Vol%
0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
0 to 5 Vol% (kPa)
0 to 6 Vol% (kPa)
0 to 5 Vol% (kPa)
0 to 10 Vol% (kPa)
0 to 20 Vol% (kPa)
Pressure-Volume
Flow-Volume
100 to 240 V~ 50/60 Hz
100 to 127 V~ 50/60 Hz
Or
220 to 240 V~ 50/60 Hz
Instructions for use Perseus A500 SW 2.0n