Dräger Perseus A500 Instructions For Use Manual page 236

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data
Measuring systems and displays (continued)
Detection
Primary gas
Secondary gas
Minimum displayed concentration
xMAC
Range
Accuracy
Resolution of displayed value
Measurement of supply pressures
Central supply
Range
Accuracy
Resolution of displayed value
Gas cylinders
Range
236
Automatic
At the latest at 0.3 Vol%
At the latest at 0.4 Vol% or 0.1 xMAC (the larger
value applies)
With a desflurane concentration greater than
4 Vol%, mixture detection occurs at the latest
when the concentration of the second anes-
thetic gas rises above 10 % of the desflurane
concentration.
The secondary gas becomes the primary gas
when the expiratory xMAC value is more than
0.2 xMAC above that of the primary gas.
The specified detection thresholds refer to rising
anesthetic gas concentrations (e.g., at the start
of an operation). If the anesthetic gas concen-
tration falls, a concentration of down to
0.05 Vol% will be measured, based on the last
anesthetic agent detected. Below this concen-
tration, a value of 0 Vol% will be displayed.
Based on patient age, anesthetic agent, and
nitrous oxide concentration
(the xMAC is corrected for ambient pressure)
0 to 9.9
Corresponds to that of the gas measurement
0.1
0 to 9.8 kPa x 100
0 to 140 psi
0 to 0.98 MPa
±4 % or ±0.2 kPa x 100 (the larger value
applies)
±4 % or ±3 psi (the larger value applies)
±4 % or ±0.02 MPa (the larger value applies)
0.1 kPa x 100
1 psi
0.01 MPa
0 to 250 kPa x 100
0 to 3600 psi
0 to 25 MPa
Instructions for use Perseus A500 SW 2.0n