WARNING
Risk of electric shock
Only connect devices to the serial interfaces,
or devices and networks to the network
interface, which have a maximum nominal
voltage of 24 Vdc and meet one of the
following standards:
– IEC 60950-1 / IEC 62368-1: ungrounded
SELV circuits
– IEC 60601-1 (as of 2nd edition): touchable
secondary circuits
WARNING
Risk of electric shock
Connecting devices to the serial data
interfaces can lead to an increased leakage
current. If the protective ground connector of
one of these devices fails, the patient leakage
current may rise above the permissible
values.
– Have the leakage current checked by
service personnel.
– If the permissible value is exceeded,
disconnect the devices from the serial
interfaces.
– Do not touch the connectors of the
interfaces and the patient simultaneously.
WARNING
Risk due to impermissible data connections
The impermissible use of data interfaces can
result in new hazards.
– Only make connections to data interfaces
with permission from the responsible
organization (IT representative and the
hospital equipment officer).
– Follow the information in the chapter
"Connections to IT networks".
Instructions for use Perseus A500 SW 2.0n
Establishing a data connection
 Connect the device to a network or a computer.
See page 27 for a picture of the interfaces.
Only use the cables from the list of accessories.
For further information on configuring the
respective interface, see page 168.