Mounting the upper part of the breathing
system housing
I
H
1 Place the upper part (G) of the breathing
system on the lower part (H).
2 Pay attention to the correct alignment of the
three levers (I), see illustration above.
3 Turn the 3 levers (I) by about 120° clockwise.
Instructions for use Perseus A500 SW 2.0n
I
G
I
Inserting the flow sensors and the ports
WARNING
Risk of fire
High temperatures arise in the flow sensors,
for example, in operation or while calibrating
during a system test. Due to the high
temperatures, residual vapors of flammable
disinfectants (e.g., alcohols) and residues
that were not removed during reprocessing
may ignite.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling, such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
K
J
M
1 Insert flow sensors (J) and (L).
2 Insert the inspiratory port (K) and screw it tight
with the knurled nut.
3 Insert in the expiratory port (M) and screw it
tight with the knurled nut.
Assembly and preparation
L
65