Dräger Perseus A500 Instructions For Use Manual page 235

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Measuring systems and displays (continued)
Drift
O
2
Range
Accuracy
Resolution of displayed value
T
10...90
CO
2
Range
Accuracy
Resolution of displayed value
T
10...90
N
O
2
Range
Accuracy
Resolution of displayed value
T
10...90
Anesthetic gases
Range
Halothane
Isoflurane
Enflurane
Sevoflurane
Desflurane
Accuracy
Resolution of displayed value
T
10...90
Instructions for use Perseus A500 SW 2.0n
Compensated by automated cyclic zeroing
Ambient air is fed into the breathing system
during automatic zeroing
0 to 100 Vol%
±(2.5 Vol% + 2.5 % relative)
1 Vol%
<500 ms
0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
±(0.43 Vol% + 8 % relative)
±(0.43 kPa + 8 % relative)
±(3.3 mmHg + 8 % relative)
0.1 Vol%
0.1 kPa
1 mmHg
<350 ms
0 to 100 Vol%
±(2 Vol% + 8 % relative)
1 Vol%
<500 ms
0 to 8.5 Vol% (kPa)
0 to 8.5 Vol% (kPa)
0 to 10 Vol% (kPa)
0 to 10 Vol% (kPa)
0 to 20 Vol% (kPa)
±(0.2 Vol% +15 % relative)
±(0.2 kPa +15 % relative)
0.1 Vol% (kPa) for desflurane
0.01 Vol% (kPa) for all other anesthetic gases
<500 ms
Technical data
235