Assembly and preparation
B1
Example of a permissible configuration:
Maximum
Arm weight
length
400 mm
4 kg
(15.7 in)
(8.8 lbs)
300 mm
3 kg
(11.8 in)
(6.6 lbs)
P1= 400 mm * (4 kg + 5 kg) = 3600 mm*kg
(P1=15.7 in * (8.8 lbs + 11 lbs) = 310.9 in*lbs)
P2= 300 mm * (3 kg + 12 kg) = 4500 mm*kg
(P2= 11.8 in * (6.6 lbs + 26.5 lbs) = 390.6 in*lbs)
3600 mm*kg + 4500 mm*kg = 8100 mm*kg
(310.9 in*lbs + 390.6 in*lbs = 701.5 in*lbs)
8100 mm*kg < 8500 mm*kg
(701.5 in*lbs < 738 in*lbs)
If a combination of arms on one device side
exceeds the value of 8500 mm*kg (738 in*lbs),
tipping stability as per IEC 60601-1 is no longer
given. Check the tilting stability.
50
B2
A
Weight of the
attached com-
ponent
5 kg
(11 lbs)
12 kg
(26.5 lbs)

Mounting on the mounting rails

Depending on the position of the mounting rail, the
following weights are permissible:
Position
On the side of the work-
ing surface
At the top on the rear of
the device
WARNING
Risk of tipping over
If the weight of the mounted accessories
exceeds the permissible maximum weight,
the medical device may tip over.
Observe the maximum weight per arm.
WARNING
Risk of tipping over
If the weight of the accessories is unevenly
distributed about the device, the medical
device may tip over.
Distribute the weight evenly.
Special features of the ceiling-mounted
version
WARNING
Risk of tipping over
If the ceiling-mounted version has been
placed on the flexibility trolley or the service
cart, the information for fitting the
accessories still applies as for the ceiling-
mounted version.
– Pay attention to the maximum weight of
the accessories.
– Pay attention to the distribution of the
weight.
Instructions for use Perseus A500 SW 2.0n
Maximum weight
10 kg
(22 lbs)
2.5 kg
(5.5 lbs)