System Setup > System - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
System setup > System
Vertical tab "General"
Headline/
Parameter
Language
Time source
IP address
Date and time
Automatic switch to
daylight savings
time
OR working hours
Device name
Configuration pass-
word
Reset all pages to
164
Setting range
List of available languages
English (United States)
MEDIBUS 1; MEDIBUS 2; NTP server; None Select source for time synchroni-
4 numeric fields
day; month; year
hour; minute
On; Off
Hour : Minute to Hour : Minute
6:30 to 18:30
Rename
Device name (up to 16 alphanumeric charac-
ters)
Perseus
Change password
Factory defaults
Description
Select language.
A flag symbol identifies the tabs
that lead to the page with the lan-
guage settings.
zation.
Prerequisite: Connected device
supports this function.
Enter the IP address of the NTP
server and confirm.
Set the date and time. The change
is applied on leaving the "General"
vertical tab.
Activate or deactivate automatic
switching to daylight saving time.
Set the working hours of the oper-
ating room.
During this time, the gas measure-
ment is kept in a preheated and
calibrated state so measured val-
ues are available after only a short
waiting period. However, this
decreases the life span of the
patient gas measurement module.
Change system name to, e.g.,
enter the installation site.
Change the configuration pass-
word.
Reset all the settings for the Sys-
tem setup dialog window to the
factory settings.
Instructions for use Perseus A500 SW 2.0n