Dräger Perseus A500 Instructions For Use Manual page 176

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
O
uptake trend
2
WARNING
Risk of incorrect therapy settings
The numerical value displayed for the O
uptake parameter is not sufficiently accurate
to enable therapy decisions to be made.
Do not make therapy decisions based solely
on the displayed numerical value. Only the
trend curve can be used for therapeutic
decisions.
O2 uptake
Primary agent
Iso
in
RR
RR
176
Econometer (optional)
2
Low-flow wizard (optional)
Gases in/et
et
Gas supply
Econometer
Deficit
Efficient
Required FG flow
Total flow
Efficient
Gases in/et
in
O2
N2O
Iso
Gas supply
O2
Air
Instructions for use Perseus A500 SW 2.0n
Surplus
et
N2O