Dräger Perseus A500 Instructions For Use Manual page 155

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Vertical tab "Alarm config. 1"
Headline/
Parameter
General alarm behavior
Open "Quick setup"
if alarm occurs
"Second agent
detected" alarm
"xMAC low" alarm
"FiO
too high for
2
neonates" thresh-
old value [%]
"FiO
too high for
2
neonates" alarm
after [h:mm]
Priority of "Apnea
ventilation" alarm
Alarm behavior in "Pause" mode
Priority of "Pause
time expired" alarm
Default value for
"Timer" [mm:ss]
Instructions for use Perseus A500 SW 2.0n
Setting range
On; Off
On; Off
On; Off
50; 25 to 90
Off; 0:10 to 9:50
0:15
Medium; Low
High; Medium; Low
0:30 to 02:00
0:30 to 02:00
2:00
1:00
Description
Opens the Quick setup window
automatically in the event of an
alarm.
Issues an alarm when an anes-
thetic gas mixture is detected.
Activate the xMAC low alarm.
Set the FiO
ered critical for neonates. If this
value is exceeded for a certain
time, the FiO
nates alarm will be triggered.
Set the time after which the FiO
too high for neonates alarm will be
triggered.
Specify the alarm priority when the
set minimum respiratory rate is not
reached in PSV ventilation mode.
Specify the alarm priority for the
alarm that is issued when the time
set in the Pause mode has
expired.
0:30 to 02:00
Specify default time for Pause.
0:30
Configuration
value that is consid-
2
too high for neo-
2
2
155