CAUTION
Risk of injury or material damage
If an accessory is fitted to Perseus which projects
below the lowest point of the anesthesia
workstation, the accessory may strike objects or
persons when the ceiling supply unit is lowered.
Only fit accessories that maintain a distance of at
least 12 cm (4.7 in) from the floor when the
anesthesia workstation is lowered to its lowest
point.
The total weight of all accessory parts that are
fastened to standard rails or are placed on the
shelves must not exceed 20 kg (44 lbs).
Additionally, all product labels regarding the
maximum load that are attached to standard rails,
shelves, and ceiling supply units must be
observed.
CAUTION
Risk of damage to the device
If the maximum load capacity of the ceiling
supply unit is exceeded, the load cannot be
supported.
Refer to the instructions for use of the ceiling
supply unit and observe the maximum load
capacity.
WARNING
Risk of injury or material damage
If Perseus is not properly mounted on the
flexibility trolley or is being dismounted from
it, the trolley may tip over or Perseus may fall
off.
The mechanical coupling and uncoupling of
Perseus and the flexibility trolley must only
be performed by specialized service
personnel.
Instructions for use Perseus A500 SW 2.0n
WARNING
Risk of injury or material damage
If Perseus is mounted on a flexibility trolley
and additional devices have been attached, it
is possible that the maximum load capacity
will be exceeded when Perseus is coupled to
the ceiling supply unit again.
– Observe the maximum load capacity of the
ceiling supply unit.
– Use only those combinations of patient
monitoring and accessories that are
permissible for the ceiling-mounted
version of Perseus.
– Observe the maximum weights and tilting
moments.