Dräger Perseus A500 Instructions For Use Manual page 159

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Headline/
Parameter
Start settings for APRV (optional)
Phigh
[mbar]; [hPa];
[cmH
O]
2
Plow
[mbar]; [hPa];
[cmH
O]
2
Tlow
[s]
Thigh
[s]
Slope
[s]
Vertical tab "Proced. 1"
Headline/
Parameter
General settings
Default maneuver
Reminder
[min]
Layout
Displayed parame-
ter
Instructions for use Perseus A500 SW 2.0n
Setting range
3 to 80
3 to 80
15
15
Off;
Off;
2 to 35
2 to 35
3
3
0.2 to 10
0.2 to 10
4.0
2.0
0.2 to 10
0.2 to 10
2.0
1.0
0 to 2
0 to 2
0.0
0.0
Setting range
One-step recruitment; Multi-step recruitment;
Insp./Exp. hold
Off, 10 to 180 Off, 10 to 180 Off, 10 to 180 If the value is not set to Variable , a
Cdyn; VT
Description
3 to 80
15
Off;
2 to 35
3
0.2 to 10
1.0
0.2 to 10
1.0
0 to 2
0.0
Defines the default maneuver.
reminder for a maneuver is given
after the first switch to a controlled
ventilation mode. Defines the time
after which, following the end of a
One-step recruitment or Multi-step
recruitment maneuver, a reminder
for a further maneuver is given.
Defines which additional parame-
ter is displayed in the One-step
recruitment and Multi-step recruit-
ment dialogs.
Configuration
Description
159