System Setup > Licenses/Options - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Headline/
Parameter
Enable pause mode
(deactivates moni-
toring mode)
System setup > Licenses/Options
Vertical tab "Licenses/Options"
Headline/
Parameter
Licenses for software options
Gas supply
Disable N
O
2
Instructions for use Perseus A500 SW 2.0n
Setting range
On; Off
Setting range
On; Off
Configuration
Description
Select whether Pause mode or
Monitoring mode should be avail-
able.
On: Pause mode is available. The
mode has an adjustable timer,
after which an alarm is generated.
Pause can be activated from
standby and from all ventilation
modes.
Off: Monitoring mode is available.
Monitoring can be activated from
Standby mode and from Man/Spon
mode.
Description
– Overview of available and
active software options.
– Activating software options.
For more information, see
page 172
A device with connectors for
nitrous oxide can be configured so
that nitrous oxide is no longer dis-
played, nor can it be selected as a
carrier gas.
Prerequisite:
– Current carrier gas is Air
– Nitrous oxide is not connected
or available
To make the settings, restart Per-
seus if necessary.
On: Nitrous oxide cannot be deliv-
ered.
Off: Nitrous oxide can be deliv-
ered.