Fresh-Gas Delivery (Mechanically Controlled Gas Mixer); Ventilator - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data

Fresh-gas delivery (mechanically controlled gas mixer)

All data are standardized to STAPD conditions.
O
concentration
2
Setting range for fresh-gas flow
Electronic measurement of fresh-gas flow
Range
Accuracy
Resolution of the value displayed on the
screen
Resolution of the value displayed in the sta-
tus display
Total flow tube
Range
Accuracy
Resolution of displayed value

Ventilator

Electronically driven ventilator, fresh-gas decou-
pled
Time-based settings:
Respiratory rate RR
Minimum respiratory rate in PSV RRmin
mode
Inspiratory time Ti
Maximum inspiratory time for supported
breaths (fixed setting)
Patient category "adult"
Patient categories "pediatric patient" and
"neonate"
226
21 to 100 Vol%
0 to at least 12 L/min (O
0 to 15 L/min (O
, Air, and N
2
±10 % or ±0.12 L/min (the larger value applies)
0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from 0.2 to 0.5 L/min)
0.05 L/min (from 0.5 to 1 L/min)
0.1 L/min (from 1 to 15 L/min)
0.1 L/min
All data are standardized to STPD conditions.
0 to 10 L/min
±10 % of the set value at 100 % O
flows >1 L/min
0.5 L/min
3 to 100 /min
±10 % of the set value or ±1/min (the larger
value applies)
Off, 3 to 25 /min
±10 % of the set value or ±1/min (the larger
value applies)
0.2 to 10 s
±10 % of the set value or ±100 ms (the larger
value applies)
4 s or 1 / (2 x RRmin) (the smaller value applies)
1.5 s or 1 / (2 x RRmin) (the smaller value
applies)
Instructions for use Perseus A500 SW 2.0n
, Air, and N
O)
2
2
O)
2
and for
2

Advertisement

Table of Contents
loading

Table of Contents