Troubleshooting
Priority
Alarm
!!! 255
Inspiratory halothane high
!! 255
Inspiratory halothane high
!
75
Inspiratory halothane low
!!! 255
Inspiratory isoflurane high
!! 255
Inspiratory isoflurane high
!
75
Inspiratory isoflurane low
!! 10
Inspiratory N
!!! 255
Inspiratory sevoflurane
high
!! 255
Inspiratory sevoflurane
high
!
75
Inspiratory sevoflurane low The inspiratory anesthetic
202
Cause
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
O high
Inspiratory N
2
%.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Remedy
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakages.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakages.
O exceeds 82
Check fresh-gas composition.
2
Press the O
the breathing system.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakages.
Instructions for use Perseus A500 SW 2.0n