External Fresh-Gas Outlet; Anesthetic Gas Receiving System; Measuring Systems And Displays - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

External fresh-gas outlet

Connection
Delivery
Pressure limitation

Anesthetic gas receiving system

Suction flow
Normal range
At lower end of restricted range
Maximum fresh-gas flow to prevent contaminat-
ing ambient air
For external breathing systems (normal
range)
For external breathing systems (restricted
range)
For internal breathing systems (restricted
range)

Measuring systems and displays

As Perseus uses more accurate values for inter-
nal calculation and alarming, small deviations
due to rounding may occur.
Airway pressure
Airway pressure
Plateau pressure
Positive end-expiratory pressure
Peak inspiratory pressure
Mean airway pressure
Range
Accuracy
Resolution of displayed value
Instructions for use Perseus A500 SW 2.0n
22 mm outer taper / 15 mm inner taper (ISO)
See "Fresh-gas delivery"
Not pressure-limited
32 to 50 L/min
14 L/min
16 L/min
5 L/min
7 L/min
Paw
Pplat
PEEP
PIP
Pmean
–20 to +120 hPa (cmH
±4 % of the measured value or ±2 hPa (cmH
(the larger value applies)
1 hPa (cmH
O)
2
Technical data
O)
2
2
O)
231