Dräger Perseus A500 Instructions For Use Manual page 239

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Operating characteristics (continued)
Maximum power consumption
Electrical mains connection for the ceiling-
mounted version
Ceiling-mounted version (without isola-
tion transformer)
Maximum power consumption
Power cable
Maximum length
Protective ground resistance
Operating voltage
Operating current
Power consumption
At 230 V~
Standby
Typical (without charging the internal
battery)
Maximum (with auxiliary power sock-
ets)
Maximum (ceiling-mounted version)
For 110 V~
Standby
Typical (without charging the internal
battery)
Maximum (with auxiliary power sock-
ets)
Maximum (ceiling-mounted version)
Power consumption
Standby
Typical
Maximum (trolley version)
Maximum (ceiling-mounted version)
Inrush current (trolley version)
With isolation transformer
Without isolation transformer
Instructions for use Perseus A500 SW 2.0n
12 A
100 to 240 V~ 50/60 Hz
4 A
5 m (16.4 ft)
Maximum 0.1 ohm
≥250 V
≥16 A
0.30 A
0.35 A
12 A
4 A
0.55 A
0.65 A
12 A
4 A
55 W
70 W
2.2 kW
400 W
Approx. 26 to 48 A peak
Approx. 18 to 34 A quasi-RMS
Approx. 8 to 14 A peak
Approx. 6 to 10 A quasi-RMS
Technical data
239