Configuration

Activating software options

The following software options require an
activation code to be entered, followed by
activation:
– Pressure Support
– APRV
– FiO
prediction
2
– Econometer trend
– Low-flow wizard
– Breathing sound
– Lung recruitment
– Lung recruitment reminder
– Inspiration hold / Expiration hold
– Logbook export
Trial licenses for these options are time-limited.
An activation code is linked with the serial number
of the respective device and cannot be transferred.
The activation codes can either be loaded from a
USB flash drive or entered manually.
 Open page System setup >
Licenses/Options > Licenses/Options (A).
System setup
A
E
C
D B
172
Loading the activation code from a USB
flash drive
Prerequisite: A USB flash drive with valid licenses
is connected to the USB interface.
 Touch the Load from USB button (B).
The activation codes are read and displayed in the
list (C).

Entering the activation code

1 Touch the Enter code button (D).
2 Enter the activation code and confirm with OK.
The license is displayed in the list (C).

Activating the licensed software option

The licensed software options must be activated
before they become available.
1 Select the corresponding license from the
list (C).
2 Touch the Activate button (E) and confirm.
3 After activating all desired licenses, restart
A
Perseus.
Instructions for use Perseus A500 SW 2.0n