Maneuvers (optional)

Perseus has various maneuvers for lung
recruitment. During a maneuver, various data
concerning the lung mechanics are displayed so
that the user can assess the progress of the
maneuver.
The reminder function (option) reminds the user to
perform a maneuver. A reminder is issued after the
first change to a ventilation mode with medium or
high breathing support and also at settable
intervals after the ending of a maneuver. The
Consider recruitment message is displayed in
the waveforms for flow and pressure.
The use of lung recruitment maneuvers is the sole
responsibility of the user.
Dräger recommends that the patient's
hemodynamics always be monitored while the
maneuvers are being performed.
Available maneuvers
– Insp./Exp. hold
– One-step recruitment
– Multi-step recruitment
1 Open the Procedures dialog window.
2 Touch the tab for the required maneuver.
Procedures
B
Relevant waveforms, trends, or measured values
are displayed in area (A) of the dialog window. The
parameters for the particular maneuver are
displayed and set in area (B).
Instructions for use Perseus A500 SW 2.0n

Inspiration hold, expiration hold

Perseus provides functions whereby a breath can
be initiated or extended, or expiration can be
extended.
This can be useful in situations where the lungs of
the patient are not supposed to move, e.g., during
use of imaging techniques.
Manual inspiration/inspiration hold
This maneuver is available in the volume-
controlled modes, the pressure-controlled modes,
and in the PSV mode, and offers the following
options:
– In the expiratory phase between 2 mandatory
A
– A mandatory breath can be extended.
Manually triggering a breath
 Touch the Man. insp./Insp. hold button (C)
Manually extending a breath
 Touch the Man. insp./Insp. hold button (C)
Perseus will trigger an extended breath or will
extend an already triggered automatic breath.
Procedures
breaths, a breath can be manually triggered
and held. The ventilation pattern of the
manually triggered breath corresponds to the
ventilation pattern for the active ventilation
mode.
briefly.
and keep it pressed for the desired time.
Operation
C
D