Checking the device configuration

Perseus can be customized to suit the
requirements of the user. Settings for the following
features are possible:
– Start settings for the ventilation

Checking the operational readiness

On the Standby page, the readiness for operation
resulting from the system test is indicated by color
(A).
A
C
Standby
Color
Meaning
Green
System is fully operational.
Yellow
System is operational with limita-
tions.
There are functional restrictions.
Take further measures to ensure
patient safety (e.g., external moni-
toring).
Red
System is not operational.
Call service personnel.
Gray
The system has not been tested.
Instructions for use Perseus A500 SW 2.0n
B
D
E
– Alarm limits
– General device behavior
For further information about the configuration, see
page148.
If the system is not fully operational, the most
important irregularities (C) are displayed and a
specific test (D) is recommended as a remedy
where possible.
Additionally, the current system leakage is
displayed in area (C).
To view details regarding the status of the system,
touch the Details... button (B) or the
Tests... button (E), see page 121.
Dräger recommends performing the system test
every 24 hours. Otherwise, it will not be possible to
ensure that the device is functional.