Repair - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Service
Component
Pressure reducer
(optional)
Cylinder pressure
reducer, type MP:
– complete
Cylinder pressure
reducer, type MK:
– Sealing
– Sieve insert
– Sintered metal filter
on the cylinder con-
nection
– Diaphragm unit
– Valve cone
– Blow-off valve
– Safety valve spring
It is recommended that only original parts from
Dräger are used and that the parts are replaced by
Dräger.

Repair

Repairs may be performed only by specialized
service personnel.
It is recommended that only original parts from
Dräger are used and that the parts are replaced by
Dräger.
220
Interval
Every 6 years
Every 3 years
Every 3 years
Every 3 years
Every 6 years
Every 6 years
Every 6 years
Every 6 years
Measure
Replace
Replace
Instructions for use Perseus A500 SW 2.0n
User group
Service personnel
Specialized service
personnel