Perseus as a ceiling-mounted version

Prerequisites and notes

Docking and undocking
Perseus may only be docked to a ceiling supply
unit by service personnel. When docking and
undocking, observe the instructions for use of the
ceiling supply unit and the service cart.
WARNING
Risk of injury or damage to the device
Safety functions on the Dräger ceiling supply
unit can be temporarily disabled with the
keyswitch.
After the device has been docked or
undocked by service personnel, make sure
that the key is not left in the keyswitch but
instead is kept in a secure place.
Establishing the cable connection between
Perseus and the ceiling supply unit
WARNING
Risk of injury
If the cable connection between Perseus and
the ceiling supply unit is not established, the
safety sensor will be inoperative.
Before using Perseus, make sure that the
cable connection has been established.
 Establish the cable connection between
Perseus and the ceiling supply unit.
 Observe the instructions for use of the ceiling
supply unit.
Instructions for use Perseus A500 SW 2.0n
Combination with Dräger ceiling supply
units
Keyswitch
The ceiling supply unit must be set up for operation
with Perseus. Located on the media column there
is a keyswitch with the following setting positions:
A Operation of Perseus or Atlan
B Docking and undocking of Perseus or Atlan
C Docking, undocking, and operation of other
Dräger devices
The keyswitch can be marked with 2 different
product labels.
Product label 1:
C