Configuration
Vertical tab "Patient"
Headline/
Parameter
Default selection for
"Start" dialog
Weight
[kg]
[lbs]
Height
[cm]
[in]
Age [years, Neo:
months]
Vertical tab "General"
Headline/
Parameter
Pinsp changes with
PEEP
Ti changes with RR
(I:E ratio is locked)
Breathing system
warmer
Auto wake-up
162
Setting range
Continue case;
New adult; New pediatric; New neonate
30 to 300
5 to 50
80
25.0
67 to 661
12 to 110
176
55
120 to 300
50 to 300
185
100
48 to 118
20 to 118
73
39
12 to 130
0 to 16
32
8
Setting range
On; Off
On; Off
On; Off
On; Off
Description
Load patient data: The button that
is selected here is preselected the
first time the Start dialog window is
opened or after changes to the
system setup.
Specify start settings for:
0.4 to 10
– Patient weight
3.0
– Patient height
0.9 to 22
6.6
– Patient age
If the Height parameter is config-
ured to Off for the New neonate
Off; 20 to 80
patient category, the correspond-
Off
ing setting control will not be dis-
Off; 8 to 31
played in the Start dialog window.
Off
0 to 24
6
Changing PEEP automatically
changes Pinsp, so that the differ-
ence between PEEP and Pinsp
remains constant.
Changing the respiratory rate
automatically changes Ti, so that
the I:E ratio remains constant.
Switch breathing system warmer
on or off.
The Start dialog window opens
automatically when ventilation
activity is detected (e.g., by
squeezing the breathing bag sev-
eral times).
Instructions for use Perseus A500 SW 2.0n
Description