Problems With Cylinder Pressure Reducers; Problems With The Vapor View Option; Support Request - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Troubleshooting

Problems with cylinder pressure reducers

Fault
The connection between gas cyl-
inder and pressure reducer leaks.
The outlet pressure rises; the
relief valve relieves the outlet of
the pressure reducer.
Leakage in the housing area

Problems with the Vapor View option

Fault
Dräger Vapor 3000/ D-
Vapor 3000 is not detected.
Control dial setting position or fill-
ing level of a Dräger Vapor 3000/
D-Vapor 3000 is not detected.
Dräger Vapor 3000/ D-
Vapor 3000 is not illuminated.
Illumination is turned on, even
though no Dräger Vapor 3000/ D-
Vapor 3000 is fitted.

Support request

If the device is configured for remote maintenance,
device information can be sent to Dräger in the
event of a problem. Proceed as follows to send a
support request to Dräger:
190
Cause
The sealing ring is damaged.
The valve seat is soiled or dam-
aged.
The diaphragm is faulty.
Cause
Vaporizer is not correctly
mounted and locked.
Sensor unit is faulty.
Sensor unit or vaporizer is soiled. Clean sensor unit and remove
Sensor unit or vaporizer is dam-
aged.
Control dial is in position T.
Illumination unit is soiled.
Illumination unit is faulty.
Vapor View option is faulty.
1 In the Standby mode, touch the Tests...
button.
2 Touch the Request support button.
Remedy
Replace the sealing ring.
Close the gas cylinder valve.
Repair by service personnel
Repair by service personnel
Remedy
Correctly mount and lock vapor-
izer.
Have item repaired by service
personnel.
foreign matter.
Have item repaired by service
personnel.
Turn control dial to position 0.
Clean illumination unit and
remove foreign matter.
Have item repaired by service
personnel.
Have item repaired by service
personnel.
Instructions for use Perseus A500 SW 2.0n