Transferring device configurations

Configurations can be exported and imported with
the aid of a USB flash drive. Configurations from
one device version (electronically or mechanically
controlled gas mixer) can be transferred to other
Perseus devices of the same version in this way.
Prerequisite: The USB flash drive is connected to
the USB interface.
 Open System setup > Im/Export config. >
Im/Export config. (A).
System setup
A
The configurations saved on the USB flash drive
are displayed in a list (B). If not all of the
configurations can be seen, delete all
configurations from the USB flash drive that are
not needed or move them to a subdirectory on the
flash drive.
The following settings are neither imported nor
exported:
– Device name
– Date and time
– IP address
Instructions for use Perseus A500 SW 2.0n
C
D
B

Importing the configuration

1 Touch one of the configurations in the list (B).
2 Touch the Import button (C) and confirm.
3 Restart Perseus.

Exporting the configuration

 To export configurations, touch the Export
button (D) and confirm.