Activation of alarms after breath detection
If no breaths have been detected even after
leaving the Standby or Pause modes, the
breathing gas is monitored with regard to an O
concentration that is too low or an anesthetic gas
concentration that is too high, regardless of the
respiratory phase. At the same time, the message
Waiting for respiratory phases is displayed in
Intelligent alarm behavior
Combined alarms
If multiple alarms occurring at the same time are
caused by the same problem, they are combined
into one alarm.
Problem
Several causes of apnea are
present.
Faults in multiple components.
This causes failure of a system
function.
Suppressed alarms
Some low-priority alarms indicate a fault in a
measurement function. If this measurement
function monitors physiological parameters, alarms
based on these parameters will not be generated.
Example:
Fault
CO
measurement is faulty
2
Instructions for use Perseus A500 SW 2.0n
the CO
temporarily inactive" is displayed beside the
relevant parameters.
2
Once 2 breaths have been detected, the message
and the symbol disappear. Only then, the
respiratory-phase-dependent O
anesthetic gas alarms are completely activated.
Alarms occurring at the same
time
Apnea (no flow)
Apnea (no pressure)
Apnea (no CO
)
2
Example:
Insp. press. sensor failure
Exp. press. sensor failure
Displayed alarm
Sample line occluded
waveform, and the symbol "Alarm
2
Combined alarm
Apnea
Ventilator failure
Non-generated alarm
Apnea (no CO
Alarms
, CO
, N
O, and
2
2
2
)
2
147