Alarms

Activating the alarms related to volume

The upper alarm limit for MV is deactivated at the
start of the ventilation, but can be set during the
ventilation. The upper alarm limit for VTi is
automatically set to 130 % of the set VT in volume-
controlled ventilation modes. In pressure-
controlled ventilation modes, it is deactivated at
the start of the ventilation, but can be set during
the ventilation.
The MV low alarm is delayed in certain cases and
is issued as follows:
– No sooner than 90 seconds after a case starts,
– No sooner than 60 seconds after changing to a
mode with greater respiratory support, see
page 259.

Alarm delay and alarm escalation

To prevent unnecessary alarms, some alarms are
not displayed immediately after a limit violation, but
after a delay. In addition, certain circumstances
can cause the alarm priority to change.
Alarm
inCO
high
2
etCO
high
2
etCO
low
2
FiO
high
2
Inspiratory N
O high
2
inAgent low
FiO
low
2
144
– No sooner than 60 seconds after an Apnea
In volume-controlled ventilation modes, the alarm
limit for the inspiratory tidal volume is automatically
set to 130 % of the set tidal volume.
Resetting the Apnea (no CO
When changing to a ventilation mode with higher
respiratory support, the Apnea (no CO
reset. If the apnea situation persists, an alarm
appears after the time specified in the table "Alarm
delay and alarm escalation".
Low
---
After two successive
respiratory phases
and 15 seconds
---
(no flow) or Apnea (no pressure) alarm.
Priority
Medium
After two successive
respiratory phases
and 15 seconds
---
---
Instructions for use Perseus A500 SW 2.0n
) alarm
2
) alarm is
2
High
---
---
After two successive
respiratory phases
and 15 seconds
or
after 30 seconds if no
respiratory phases are
detected