Dräger Perseus A500 Instructions For Use Manual page 156

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
Vertical tab "Alarm config. 2"
Headline/
Parameter
Deactivate the alarm limit when activating Man/Spon, CPAP,
CPAP/PSV with ∆Psupp < 5, or Pause?
FiO
high
2
MV low
MV high
xMAC low
etCO
low
2
etCO
high
2
inCO
high
2
inAgent low
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
high
2
MV low
MV high
inAgent low
Vertical tab "Alarm config. 3"
Headline/
Parameter
Alarm limits for "Cylinder almost empty"
O
2
Air
N
O
2
156
Setting range
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Setting range
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
[bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290
Description
The CPAP and CPAP / PSV
modes are optional.
Yes; No
Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only to a
change to a ventilation mode with
lower or no breathing support (see
page 259).
The alarm behavior at the start of
the therapy is defined by the con-
figuration in the vertical tab Alarm
limits.
Specify alarm behavior in CBM
mode.
Description
Specify alarm limits for supply
pressure of connected gas cylin-
ders.
Instructions for use Perseus A500 SW 2.0n