Dräger Perseus A500 Instructions For Use Manual page 161

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Headline/
Parameter
Pressure decrease
per step when
PEEP ≤ 15
[mbar]; [hPa];
[cmH
O]
2
Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Headline/
Parameter
Start settings for fresh gas
FG O
2
[%]
FG flow
[L/min]
Minimal O
flow
2
(carrier gas: Air)
[mL/min]
Minimal O
flow
2
(carrier gas: N
O)
2
[mL/min]
Carrier gas
Instructions for use Perseus A500 SW 2.0n
Setting range
1 to 10
1 to 10
2
2
Setting range
21 to 100
21 to 100
100
100
0.20 to 15.00
0.20 to 15.00
2.00
2.00
Off, 50 to 300
Off, 50 to 300
Off
Off
50 to 300
50 to 300
200
200
Air; N
O
Air; N
2
1 to 10
2
21 to 100
100
0.20 to 15.00
2.00
Off, 50 to 300
Off
50 to 300
200
O
Air; N
O
2
2
Configuration
Description
For PEEP ≤15: Specifies the pres-
sure by which PEEP and Pinsp will
be reduced in steps.
Description
Select start settings for the fresh-
gas delivery.
Set the O
concentration.
2
Set the flow.
Set the minimum O
flow that is
2
delivered when Air is used as car-
rier gas.
Do not set this value too small;
recommended is, e.g., 200 for
adults, 100 for pediatric patients,
and 50 for neonates.
Set the minimal O
flow that is
2
delivered when N
O is used as
2
carrier gas.
Do not set this value too small;
recommended is, e.g., 200 for
adults, 100 for pediatric patients,
and 50 for neonates.
Set the carrier gas.
161