Anaesthesia ventilation
Condensation
Check the water trap
– Regularly check the water trap on the left-hand side
of the Cicero EM. When the level comes close to the
»Full« mark, remove the jar by pulling it down and
empty it.
– Fit the jar firmly back in place!
If the permitted maximum level is exceeded, the
measured gas sampling system is switched off
automatically.
The advisory message »WATER TRAP? !« is displayed on
the system screen.
CO
, O
, breathing gas and N
2
2
then no longer be possible!
Remedy: Drain the water trap as described.
Check water separators
– Water separators must be fitted at the lowest point in
the hose and hang downwards.
– They must be checked regularly and drained if
necessary.
Note hygiene regulations -
risk of infection!
The hose system remains sealed.
– Container must be replaced securely.
Anaesthetic gas scavenging line (AGS)
– Check the transfer hose.
Condensate may have collected in the hose – drain if
necessary.