Dräger Perseus A500 Instructions For Use Manual page 160

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
Vertical tab "Proced. 2"
Headline/
Parameter
Default settings for one-step recruitment
Pressure
[mbar]; [hPa];
[cmH
O]
2
Duration
[s]
Default settings for multi-step recruitment
PEEP max
[mbar]; [hPa];
[cmH
O]
2
Pinsp max
[mbar]; [hPa];
[cmH
O]
2
∆Pressure
[mbar]; [hPa];
[cmH
O]
2
Breaths/Step
Breaths@Max
Pressure rise per
step
[mbar]; [hPa];
[cmH
O]
2
Pressure decrease
per step when
PEEP > 15
[mbar]; [hPa];
[cmH
O]
2
160
Setting range
PEEP + 1 to
PEEP + 1 to
80
80
30
25
3 to 40
3 to 40
30
15
PEEP to 35
PEEP to 35
20
15
15 to 80
15 to 80
35
30
5 to 30
5 to 30
10
10
1 to 20
1 to 20
3
4
1 to 20
1 to 20
6
8
2 to 10
2 to 10
5
4
2 to 10
2 to 10
5
4
PEEP + 1 to
Sets the pressure level for the
80
maneuver.
20
3 to 40
Sets the duration of the maneuver.
5
PEEP to 35
Sets the maximum PEEP pressure
12
for the maneuver.
15 to 80
Sets the maximum inspiratory
25
pressure for the maneuver.
5 to 30
Sets the pressure difference
10
between Pinsp and PEEP with
which pressure-controlled ventila-
tion is carried out.
If PEEP max or Pinsp max is
reached in the course of the
maneuver, the pressure difference
will be reduced in steps until the
other set value is also reached.
The smallest possible value must
be at least 3 above the set value
for the pressure rise for each level.
1 to 20
Number of breaths at a pressure
5
level during the increase or reduc-
tion
1 to 20
Number of breaths at the Pinsp
10
max pressure level
2 to 10
Specifies the pressure by which
3
PEEP and Pinsp will be increased
in steps.
2 to 10
For PEEP >15: Specifies the pres-
3
sure by which PEEP and Pinsp will
be reduced in steps.
Instructions for use Perseus A500 SW 2.0n