Measuring systems and displays (continued)
Accuracy
Resolution of displayed value
Test results
Total leakage
Range
Resolution of displayed value
Leakage in automatic ventilation
Range
Resolution of displayed value
Hose compliance
Range
Resolution of displayed value
Display of calculated values
Measurement of consumption and production
CO
production of the patient
2
Range
Accuracy
Resolution of displayed value
O
uptake of the patient
2
Range
Accuracy
Resolution of displayed value
Fresh-gas consumption
Range
Accuracy
Resolution of displayed value
Instructions for use Perseus A500 SW 2.0n
±4 % or ±6 kPa x 100 (the larger value applies)
±4 % or ±87 psi (the larger value applies)
±4 % or ±0.6 MPa (the larger value applies)
1 kPa x 100
1 psi
0.1 MPa
10 to 5000 mL/min
1 mL/min
10 to 5000 mL/min
1 mL/min
Compliance
0 to 9.9 mL/hPa (mL/cmH
0.1 mL/hPa (mL/cmH
2
0 to 9999 mL/min
±25 % or 100 mL/min (the larger value applies)
1 mL/min
0 to 9999 mL/min
±25 % or 100 mL/min (the larger value applies)
1 mL/min
0 to 99999 L, only gas delivered by the gas
mixer is measured
±15 %
1 L
Technical data
O)
2
O)
237