Dräger Perseus A500 Instructions For Use Manual page 233

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Measuring systems and displays (continued)
Low-flow wizard
Range
Accuracy
Respiratory rate
Total
Spontaneous
Mandatory
Range
Accuracy
Resolution of displayed value
T
0...90
Dynamic compliance
Compliance
Mean compliance
Range
Accuracy
Resolution of displayed value
Resistance
Range
Accuracy
Resolution of displayed value
Instructions for use Perseus A500 SW 2.0n
0 to 8 L/min
±25 % of measured value or ±100 mL/min (the
larger value applies)
RR
RRspon
RRmand
0 to 150 /min
±1 /min or 10 % (the larger value applies)
1 /min
<45 s (RR ≥6 /min)
<105 s (RR <6 /min)
Cdyn
Cdyn mean
0 to 200 mL/hPa (mL/cmH
±30 % or ±3 mL/hPa (mL/cmH
value applies)
If during automatic ventilation there is still an
inspiratory flow at the end of inspiration, the
device will determine how much volume could
not be applied. This volume will then be
included with the normal volume in the compli-
ance calculation.
With increasing spontaneous breathing activity,
the compliance values can be distorted. In this
case the measurement accuracy may be
reduced.
0.1 mL/hPa (mL/cmH
2
R
0 to 100 hPa / (L/s) (cmH
±3 hPa / (L/s) (cmH
O / (L/s)) or ±30 % of the
2
measured value (the larger value applies)
(With increasing spontaneous breathing activity,
the R values can be severely distorted. Hence
maintaining measurement accuracy cannot be
guaranteed with spontaneous breathing).
1 hPa / (L/s) (cmH
O / (L/s))
2
Technical data
O)
2
O) (the larger
2
O)
O / (L/s))
2
233