Selecting IPPV mode
After switching on, the ventilation parameters set prior to
delivery or subsequently programmed by DrägerService
are active in IPPV mode.
1 Set the values required for the patient.
Default settings upon delivery of new apparatus:
(may be altered by DrägerService if requested by
customer)
V
Tidal volume
T
f
IPPV frequency
IPPV
P
Maximum ventilation pressure
max
T
:T
Inspiration/expiration time ratio
I
E
T
:T
Ratio of inspiratory pause
IP
I
I
time/inspiration time
PEEP Positive end-expiratory pressure
2 Press
.
3 Display in dialogue field:
IPPV Mode ?
4 Press rotary control to confirm -
ventilation starts -
The parameters P
, V
max
movement are displayed on the ventilator.
The system screen is also started and the IPPV alarm
limits are activated.
Adjusting ventilation parameters (P
5 Press »P
«.
max
6 The set value appears in the window above, while
3 the set value (left), the parameter and its unit of
measure (middle) and the value to be adjusted (right)
appear in the dialogue window -
4 turn rotary control:
3 the value on the right changes -
4 continue turning until the required maximum pressure
is reached. Press rotary control to confirm this value -
6 the new value is displayed -
if the new value is not confirmed, it will not be
adopted by the system and the display disappears
after approx. 10 seconds.
f
, V
, T
:T
, T
:T
, PEEP and f
IPPV
T
I
E
IP
I
like P
after pressing the corresponding parameter
max
key.
0.6 L
12 per min.
25 mbar
1 : 1.7
10 %
0 mbar
, f
, PEEP and piston
T
IPPV
for example):
max
are adjusted
IMV
3
MAN
2
SPONT
IPPV
23
600
12
IPPV Mode ?
P
max
T
IP
:T
I
f
SIMV
PCV
V
T
f
T
:T
PEEP
IMV
IPPV
I
E
1
P
PCV
Flow
MAN
SPONT
IPPV
5
32
23 Pmax / mbar 32
6
P
max
T
:T
IP
I
f
SIMV
PCV
V
T
f
IPPV
T
I
:T
E
PEEP
IMV
P
PCV
Flow
3
Anaesthesia ventilation
IPPV
TEST
4
4
33