Basic screen configuration

Access to Configuration

1 When you press the »Config« screen key, the screen
opposite is displayed:

Default values

The default settings are permanent. They are activated
after every modification and every time the device is
switched on.
By contrast the standard alarm limits are activated after
each standby.
The standard are protected by 4-digit code against
unauthorised modification.
Dräger Service can deactivate this protective function
or can program a different code freely selectable by the
user.
Enter code:
Select the digits one by one by turning the rotary
control and press to confirm. After the fourth correct
digit has been entered, access is granted to the
selection range.
(The operating procedure for the basic screen
configuration is described on the next page).
Standby
Cicero-EM
Ventilator
Data Manager
SW-Version 7.n
SW-Version 4.n
06-12-98
06-12-98
Select ventilation mode at the ventilator or switch on screen by
Standby / Configuration
Anesth.Agent
Calibration
selection
more
man.
auto
gas
Isoflurane
The anesthetic
agent is automati-
cally selected by
agent detection
06-12-98
11:59
Check system configuration. Please confirm !
00:00
Standby / Configuration
06-12-98
11:59
00:00
System screen functions
Basic screen configuration
Setting code
leakage
test
delete
Parameterbox
trend
SW-Version 2.n
06-12-98
1
config.
1
alarms inactive !
Default values
mode
adult
neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
Default values
These default settings are password
protected.
The default settings are active
on start-up.
To change the default settings
enter the password:
0 1 2 3 4 5 6 7 8 9
* * * *