Anaesthetic Gas Scavenging System - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Preparing for use
AGS-system
Bronchial aspirator
Installing the anaesthetic gas scavenging
system:
Hook the collecting system to the mount on the back
of the Cicero:
Place the slots of the receiving system over the
holders and slide the receiving system down.
See also "What's what" on page 146, No. 14.
Connecting the transfer hoses
1 Pass transfer hose from behind through the hole in
the Cicero EM and
2 connect the hose to the scavenging connector of the
breathing system (only on first installation - the hose
then remains on the connector).
3 Connect the transfer hose to the connector on the
receiving system.
Make sure the connection is secure!
4 Make sure that the connector for the second transfer
hose is sealed with a screw plug.
Do not plug the openings of the receiving system, other-
wise the breathing system could be drained.
5 Connect the scavenging hose to the output connector
of the scavenging system.
6 Plug the connector of the scavenging hose into the
wall socket. The visual indicator shows »green«.
18
D
2
D
4
5
3
1
6
Detailed view
from below after
removing soda
lime container.
1