Operating Modes For Adults And Neonates; Alarm Limits - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Mode
The Cicero EM has separate operating modes for adults
(adult) and neonates (neon). These modes differ in the
following way:
– Default alarm limits (see below)
– Volumeter measuring range (see page 58)
– NiBP functions (see page 91)
– ECG functions (see page 89).
Every time the Cicero EM is switched on, it automatically
operates in the mode selected in this menu.

Alarm limits

The standard alarm limits are defined
– separately for adults and neonates,
and, in order to activate/deactivate the MAN/SPONT
alarm limits, for
– et CO
,
2
– in CO
,
2
– O
high and
2
– AMV.
The default alarm limits are automatically activated after
either:
– switching on the Cicero EM,
– leaving »Standby« mode,
– selecting »default« under »Alarms«,
– switching over between »Neon« and »adults« mode
The limit values are allocated in groups to the following
measurement parameters:
– iBP
– ECG / SpO
/ NiBP / Temp
2
– Anaesth. Agent
– Pressure / Ventilation
Both the upper and lower limit is displayed at the limit
symbol:
_
22
Upper limit
11
Lower limit
Two dashes (– –) instead of a number indicate that this
alarm limit is deactivated and not monitored. This is set
by turning the rotary control respectively above or below
the maximum or minimum settable value and pressing to
confirm.
Not all the alarm limits can be deactivated!
Configuring the system screen in »Standby«
Standby / Configuration
06-12-98
11:59
Select the required ventilation mode (adult or neonate). Please confirm !
00:00
Standby / Configuration
standard adult
standard neonates
Man/Spont-alarmlimits
et CO
on
2
in CO
on
2
O
high
on
2
MV
on
06-12-98
11:59
Standard alarm limits configuration menu. Please confirm !
00:00
Standby / Configuration
Standard - Adult
ECG / SpO
/
iBP
Anesth. Agent
2
NiBP / Temp
Aorta
160
120
50
80
HR
Fi Hal
sys
100
120
Pulse
Fi Iso
dia
50
50
A.Pulm
24
--
8
SpO
sys
Fi Enfl.
92
2
24
8
NIBP/ART
Fi Sevo
mean
100
sys
50
20
--
Fi Des
dia
160
dia
CVP
8
80
0
mean
--
ICP
T
--
16
1
0
mean
--
T
mmHg
160
2
--
80
sys
--
ST
24
--
8
mean
100
dia
50
06-12-98
11:59
Standard alarm limits configuration menu. Please confirm !
00:00
System screen functions
alarms inactive !
Default values
mode
adult
neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
Settings
mode
adult
neo.
alarm limits
parameter
screen
off
acoustic
off
list entry
off
off
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
Settings
Pressure /
mode
adult
neon.
Ventilation
1,5
12,0
alarm limits
MV
--
3,0
2,3
40
parameter
PAW
--
8
3,4
50
screen
etCO
2
--
--
3,4
--
acoustic
FiO
2
--
20
12,0
list entry
--
transport
RS 232 (Medibus)
record (printer)
basic config.
49