Protection
Protection Module Configuration with IMx-M Manager
Table 2-23: Valid filter selections for different measurement types.
Type
Temperature
Radial shaft vibration
Position
Complementary differential expansion
Piston rod drop (average mode)
Piston rod drop (triggered mode)
Casing vibration
Eccentricity
Other
Gap
1 If this filter is activated, then only one channel of the pair can be used.
2 This filter cannot be used if integration is selected.
2 - 116
Filter
DC
101 000 Hz,
51 000 Hz,
1600 Hz,
44 000 Hz,
1
105 000 Hz
1
1 00010 000 Hz
User defined
DC
DC
-
-
101 000 Hz
1600 Hz
2
44 000 Hz
51 000 Hz
1,2
105 000 Hz
1,2
1 00010 000 Hz
User defined
010 Hz
DC
DC
SKF Multilog On-Line System IMx-M
User Manual
Need help?